Chapter 5 _________________________________________________________________ Operation

The response is the following:

allccccccccmmmmmmvvvxxx . . . xxx<CR><LF>

where

a

=

Single-digit WMT700 address that corresponds to the
first character of the address value

11

=

SDI-12 version number, indicating SDI-12 version
compatibility; for example, version 1.1 is encoded as 11

cccccccc

=

8-character vendor identification Vaisala_

mmmmmm =

6 characters specifying the model number of WMT700

vvv

=

3 characters specifying the firmware version 604

xxx ... xxx =

Optional field, up to 13 characters, used for a serial
number or other specific WMT700 information that is
not relevant for operation of the data recorder (not used)

<CR><LF> =

Terminates the response

aM! — Start Measurement

This command starts measurement. WMT700 does not return
measurement data after this command. It sends a response that
indicates when the measurement result can be retrieved using the
aD0! command and the number of parameters.

aM!

where

a

= Single-digit WMT700 address that corresponds to the

first character of the address value

M

= Start measurement command

!

= Terminates the command

VAISALA _______________________________________________________________________ 173