Ws425 a/b nmea standard data message, Missing readings – Vaisala WMT700 User Manual

Page 157

Advertising
background image

Chapter 5 _________________________________________________________________ Operation

To start continuous measurement, use the START command.
You can stop the measurement with the STOP command. For
instructions, refer to sections START — Start Continuous
Measurement
and STOP — Stop Wind Measurement on page
115.

WS425 A/B NMEA Standard Data
Message

The standard variable length, comma-separated, MWV wind
message is defined by NMEA 0183 V2.20 as follows:

$WIMWV,<dir>,<ref>,<spd>,<uni>,<sta>*<chk><CR><LF>

where

$WIMWV = Fixed text
<dir>

= Wind angle: 0 to 359 degrees

<ref>

= Reference:

R = Relative

<spd>

= Wind speed

<uni>

= Wind speed unit:

K = Kilometers per hour
M = Meters per second
N = Knots

<sta>

= Status:

A = Data valid
V = Invalid data

*

= Fixed text

<chk>

= Checksum (8-bit XOR, excluding $ and *)

<CR>

= Carriage return code, ASCII 0DH

<LF>

= Line feed code, ASCII 0AH

NOTE

When the NMEA Standard profile is selected, the autoInt parameter
must have a non-zero value since no polling command is defined for this
profile.

Missing Readings

If data is missing due to a measurement problem, the NMEA
messages show "V" in the status field. Wind speed and wind
direction fields are left empty.

VAISALA _______________________________________________________________________ 155