Chapter 3 _______________________________________________________ Functional Description

Missing Readings and Error
Indication

If WMT700 is not able to measure the wind, it indicates a missing
reading in the output. Most common reasons for measurement
problems are foreign objects (such as ice, birds, or other foreign
objects) on the line of measurement or sound reflections from
nearby objects (such as wind tunnel walls).

The default error indication is an out-of-range signal, which is
more than 10 V or 20 mA, but other error settings can also be
configured.

Example

To set analog output 1 error indication in current mode to 2 mA,
set the analog output error value to 0.002. Enter the following
command:

S aout1err,0.002

For more information, see section Appendix D, Configuration
Parameters, on page 205.