Chapter 3 _______________________________________________________ Functional Description

NOTE

The scaling of the current output 4…20 mA depends on the sensor
measurement range (WMT701, WMT702, or WMT703), while the
0…20 mA output has 0.2 mA / m/s scaling for all the measurement
ranges.

Table 20 and Table 21 below list the most common settings for
different units.

Table 20

Common Transfer Function Settings for
AOUT1 (WS)

Output Signal

Scaling/Units

Setting for Gain

Setting for Offset

Voltage

8 mV / mph
100 mV / m/s

0.017895
0.1

0
0

Current

0.2 mA / m/s

0.0002

0

Frequency

5 Hz / mph
10 Hz / m/s

11.185
10

0

WS425 - voltage

8 mV / mph

0.017895

0

WS425 - frequency 5 Hz / mph

11.185

0

Table 21

Common Transfer Function Settings for
AOUT2 (WD)

Output Signal

Scaling/Units

Setting for Gain

Setting for Offset

Voltage

4 mV / °
20 mV / °

0.004
0.02

0
0

Current

50 uA / degree

0.00005

0

Potentiometer

359° = Aout ref 0.0027855

0

You can configure output scaling or transfer function settings in a
variety of ways by changing the custom gain and offset. The basic
measurement units are m/s and degrees. The physical output units
are V, A, and Hz. For the potentiometer, an output of 1 means
100% of the Aout ref voltage.

The following formula illustrates the impact of gain and offset
values on the produced output:

o = y0 + k×s

where

o = Produced analog output (V, A, Hz, 100%)
s = Measured wind speed or direction (in m/s or °)
k = Selected gain value
y0 = Selected offset value