Figure 40, Wiring of heated wmt700 versions, part 1, Figure 41 – Vaisala WMT700 User Manual

Page 90: Wiring of heated wmt700 versions, part 2, Caution

Advertising
background image

User's Guide ______________________________________________________________________

Figure 40 and Figure 41 below illustrate the wiring of heated
WMT700 instruments.

1301-018

Figure 40

Wiring of Heated WMT700 Versions, Part 1

Note that when using separate power supplies for operating and
heating, the minus (-) terminals of the power supplies are
connected together by an additional wire. Use a minimum
0.75 mm

2

wire for connecting the (-) terminals together.

1307-003

Figure 41

Wiring of Heated WMT700 Versions, Part 2

CAUTION

There are two terminals connected in parallel for both positive and
negative rails of the heating voltage for the maximum current capacity. In
case the connection cable has parallel supply wires, they all have to be
connected to ensure current capacity. Leaving one terminal unconnected
or connecting it to the ground may cause a WMT700 malfunction or a
short circuit in the power supply.

88 ___________________________________________________________________ M211095EN-E