The Compex Performance is designed for use in the electromagnetic

environment stipulated below. The buyer or user of the Compex

Performance must ensure it is used in this recommended environment.

RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -

NOTE: VT is the AC supply voltage before application of the test
level.

Immunity test

Test level IEC 60601

Electrostatic discharge (DES)

IEC 61000-4-2

± 6 kV at the contact

± 8 kV in air

Rapid electrical transients in

bursts of noise

IEC 61000-4-4

± 2 kV for power supply

lines

± 1 kV for input/output lines

Surge currents
IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

Voltage trough, short power

cuts and voltage variations on

power supply input lines

IEC 61000-4-11

< 5% VT (trough > 95% of

VT) for 0.5 cycle

< 40% VT (trough > 60% of

VT) for 5 cycles

< 70% VT (trough > 30% of

VT) for 25 cycles

< 5% VT (trough > 95% of

VT) for 5 seconds

Observance level

± 6 kV at the contact

± 8 kV in air

Not applicable

System battery-powered

Not applicable

System battery-powered

Not applicable

System battery-powered

Magnetic field at mains

frequency (50/60 Hz)

IEC 61000-4-8

3 A/m