English – Compex Performance User Manual

Page 313

Advertising
background image

311

English

The Compex Performance is designed for use in the electromagnetic

environment stipulated below. The buyer or user of the Compex

Performance must ensure it is used in this recommended environment.

- ELECTROMAGNETIC IMMUNITY

NOTE: VT is the AC supply voltage before application of the test
level.

Electromagnetic

environment - recommendations

Portable and mobile RF communication devices must only be used relative to the Compex

Performance and its wiring at a distance which is not less than the spacing recommended and

calculated using the appropriate equation for the transmitter’s frequency. Recommended

spacing

d = 1.2 √P

d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2.5 GHz

Where P is the maximum output power of the transmitter in watts (W) set by the manufacturer’s

specifications and where d is the recommended spacing in metres (m). The field intensity of RF fixed

transmitters, as determined by an electromagnetic survey

a

must be less than the observance level

to be found in each frequency range

b

.

Interference may occur close to any appliance identified by the following symbol:

Electromagnetic wave propagation is modified by absorption and reflection due to buildings,

objects and persons.

In the event of abnormal operation, new measures may then be imposed, such as realignment

or movement of the Compex Performance.

b

Above the frequency amplitude from 150 kHz to 80 MHz, the field intensity must be

< 3 V/m

PERFORMANCE FR DE IT ES EN:ENGY Man FR 14/02/08 15:12 Page 311

Advertising