November 2014

Software description

103

The function reserves of absolute track, incremental track and position value formation are
evaluated in bar diagrams; the result is displayed in %. A drag indicator (triangle below the bar
display) marks the measured minimum.
Green range: The output signal is within the specifications.
Yellow range: The output signal is outside the specifications, but no counting or calculation
errors are to be expected. No alarms are generated, warnings may occur.

8

Click this button to stop the recording.

8

When you click the "Encoder parameters" button, the log display appears.

Measuring range and recording period are displayed now (in green color).

Note

The yellow range indicates: Service or maintenance recommended!