1 test of forced dynamic sampling – HEIDENHAIN PWM 20 User Manual

Page 110

Advertising
background image

November 2014

Software description

111

5.6.1 Test of forced dynamic sampling

8

Click the "Next >" button to prepare the test of forced dynamic sampling. Forced dynamic
sampling means that error bits are intentionally triggered and their reactions evaluated.
If there are error messages in the status display of the encoder (alarms or warnings),
the message "Error messages are present" appears.
At this time, the error bits can be deleted by pressing "Yes", and the wizard can continue.

8

If you click "No", the wizard is aborted.

8

The supported errors are listed in a table. Press the "Start" button to show the test result.

Supported errors:

Note

Error bits that emerge later can be cleared from the screen, but are saved in the background.
In this case, the error entry will appear at the end of the test in the EnDat status of the log
display.