Checking incremental encoders

205

6.3.7 Logic function, bar graphs

The bar graphs include

Measurement of on-to-off ratios TV A / TV B and phase angle

Measurement of REF signal position and width

Measurement of minimum edge separation

Designations of the logic bar graphs:

Colors of the pointers in the bar graphs

The higher the frequency of the output signal becomes, the smaller is the edge separation.

Note

The displays "Incremental signal characteristics" and "Reference signal" are only useful for
encoders that feature non-clocked interpolation (mostly TTLx1 or TTLx2).
For higher interpolation factors, clocked interpolations (TTLx5, x50, x200 ...) are used in most
cases. The edges of the square-wave signal are falsified and do no longer correspond to the
original signal.

TV A, TV B

On-to-off ratio (Tastverhältnis) = ratio of ON signal to OFF signal
(180° / 180° = 1 : 1) in [°]

Pha

Phase angle difference of signals A and B in [°]

LR

Position error of the reference mark in [°]

BR

Width of the reference mark in [°]

FA

Minimum edge separation in [µs]

Green bars:

Signals within the specified tolerance

Red bars:

Signals outside the specified tolerance

Multiple
arrows:

Scaling exceeded

Note

The standard tolerance limits are marked by red lines in the bar graphs (see pointers).