Standard signal amplitudes and tolerance ranges – HEIDENHAIN PWM 20 User Manual

Page 156

Advertising
background image

November 2014

Checking incremental encoders

159

Standard signal
amplitudes and
tolerance ranges

Designations of the bar graphs and calculations

The following information and formulas refer to the interface description and the signal diagrams
in there (see chapter “Interface description” on page 209).

1 Vpp

11 µApp (25 µApp)

min.

typical

max.

min.

typical

max.

Sig A and Sig B

0.6 Vpp

1 Vpp

1.2 Vpp

7 µApp

(15 µApp)

11 µApp

(25 µApp)

16 µApp

(35 µApp)

A / B

0.8

1

1.25

0.8

1

1.25

Pha

80 °

90 °

100 °

80 °

90 °

100 °

TV A and TV B

- 15 °

0 °

+ 15 °

- 15 °

0 °

+ 15 °

Note

The stated tolerances are HEIDENHAIN standard values.
High-accuracy encoders (e.g. angle encoders), encoders for large temperature ranges
(e.g. motor encoders) or for high speeds may have different limit values.
Please always refer to the original documentation of the encoders to be checked (mounting
instructions).
In case of doubt, contact the HEIDENHAIN helpline (see chapter “Contacts” on page 243).

Note

You can find the standard signal amplitudes and tolerances in the interface
descriptions.
Always observe the tolerances stated in the original documentation (mounting
instructions, etc.) of the encoder to be checked!
In case of doubt, contact the HEIDENHAIN helpline (see chapter “Contacts” on
page 243)
.