Bar display of reference signal – HEIDENHAIN PWM 20 User Manual

Page 160

Advertising
background image

November 2014

Checking incremental encoders

163

Bar display of
reference signal

Display of the signal parameters and tolerances as bar graphs with tolerance markings.

Recommendation:
Traverse the reference mark(s) from both sides; on encoders with distance-coded reference
marks make random samples and check "faulty" areas.

Standard signal amplitudes and tolerance ranges of the reference signal

1 Vpp

11 µApp

min.

typical

max.

min.

typical

max.

L R

- 60

0

+ 60

- 60

0

+ 60

B R

180°

360°

540°

180°

360°

540°

R R

0.04 V

-

1.7 V

0.4 µApp

-

25 µApp

N R

0.2 V

-

0.85 V

0.2 V

-

0.85 V

S R

0.2

-

0.8

0.2

-

0.8

Note

The stated tolerances are HEIDENHAIN standard values.
High-accuracy encoders (e.g. angle encoders), encoders for large temperature ranges
(e.g. motor encoders) or for high speeds may have different limit values.
Please always refer to the original documentation of the encoders to be checked (mounting
instructions).
In case of doubt, contact the HEIDENHAIN helpline (see chapter “Contacts” on page 243).