HEIDENHAIN ATS Software User's Manual

Example: LC

1

Status display: Errors and warnings

2

Position displays absolute / incremental

3

Signal position; verification whether datum shift is in compliance with EnDat
(yellow symbol = EnDat non-compliant)

4

Different velocity ranges

5

Permitted accuracy deviations [LSB] at defined velocity ranges [m/min]

6

Velocity at deviation

7

Determined deviation spans [LSB] of the different velocity ranges

8

Reset deviation spans

Invert counting direction of incremental counter

Display details on alarms and warnings

Note

If the signal position is "yellow" (datum is EnDat non-compliant), a TNC control generates an
error message. Depending on the TNC's signal resolution an EnDat non-compliant datum
shift may result in a dimensional error that is beyond the accuracy specifications of the
machine.