Checking incremental encoders

199

Encoder
characteristics

6.3.3 Level function

Bar of oscilloscope settings, TTL

Those function keys are described that are new or differ from those of the analog output signals.

Position:

The counter counts signal periods

Frequency:

Current input frequency

Failure signal
(-UaS):

"Left" LED =

Concurrent failure signal (red color only as long as
an error is present)"Right“ LED =

Failure signal is logged (display permanently red,
if an error was detected within the measuring range)

Note

The designations, scalings and units are automatically adapted to the connected encoder.
The meanings of the diagrams, curves and units can be seen from context menus; simply
move the cursor to the desired position, and the menu will open (see illustration below).

Reference trigger on/off: The reference signal is shown through a trigger LED.
The graphics display 3 shows the triggered, "frozen" reference pulse.

Note

This function has to be activated before ref. mark level measurement (bar display Ua0 H) is
possible! To make the reference mark known, the encoder must traverse it first.

TTL level measurement, positive/negative signals: Switch between inverted