3 test by comparison of position values 1 and 2 – HEIDENHAIN PWM 20 User Manual

Page 115

Advertising
background image

116

HEIDENHAIN ATS Software User's Manual

5.6.3 Test by comparison of position values 1 and 2

For this test the position value 1 is compared to the position value 2. The values must not differ
by more than 1 bit.

8

Press the "Measurement" button to start the comparison.
Traverse the largest possible range until all edges are deleted from the "Collected positions"
diagram (line at 1).

Note

Traverse the largest possible measuring range!

Advertising