Hamilton Electronics EPHUM011 User Manual

Page 12

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 12 / 67

2.4.5.3

Reading the Minimum and Maximum Possible Physical Output Current

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

4366

4

Min physical
output current
for AO1 [mA]

Max physical
output current
for AO1 [mA]

3, 4

U/A/S

none

4494

4

Min physical
output current
for AO2 [mA]

Max physical
output current
for AO2 [mA]

3, 4

U/A/S

none

Figure 2.4.5.3.1: Definition of register 4366 / 4494

Command: Limits AO1

Modbus address:

4366

Length:

4

Type:

3

Read

Parameter:

Min limit [mA]

Max limit [mA]

Format:

Float

Float

Value:

3.5

22

Figure 2.4.5.3.2: Example to read the min and max output current of AO1. Min is fixed to 3.5 and Max

is fixed to 22 mA (Currents above 20 and below 4 mA indicate erroneous measurements or
errors).
The same values are stored in register 4494 for AO2.

2.4.5.4

Reading the Minimum, Maximum and Mid Current for Measurement Value Output

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus
function
code

Read
access

Write
access

4370

6

Min output for
measurement
value for AO1
[mA]

Max output for
measurement
values for AO1
[mA]

Mid output
(bilinear) for
measurement
values for
AO1 [mA]

3, 4

U/A/S

none

4498

6

Min output for
measurement
value for AO2
[mA]

Max output for
measurement
values for AO2
[mA]

Mid output
(bilinear) for
measurement
values for
AO2 [mA]

3, 4

U/A/S

none

Figure 2.4.5.4.1: Definition of register 4370 / 4498

Command: MinMaxMid current AO1

Modbus address:

4370

Length:

6

Type:

3

Read

Parameter:

Min current [mA]

Max current [mA]

Mid current [mA]

Format:

Float

Float

Float

Value:

4

20

12

Figure 2.4.5.4.2: Example to read the min, max and mid output current for measurement values for

AO1. They are fixed to 4, 20 and 12 mA.

The same values are stored in register 4498 for AO2.

Note:
Mid current must always be defined. However, in linear output mode, the mid current value has no
physical meaning and will not affect the 4-20 mA output.