Hamilton Electronics EPHUM011 User Manual

Page 36

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 36 / 67

Standard Calibration with Automatic Recognition of the Standard

Prior to calibration, the specialist selects - for each set of calibration standards - a list of calibration
standards allowed for automatic recognition (register 9530).
When the calibration is initiated, the sensor screens this list of allowed calibration standards and
checks if the currently measured electrical potential is within the allowed range of electrical potentials,
defined for the individual calibration standards in this list.

If the sensor does find a corresponding calibration standard, the nominal pH value of the specific
calibration standard is assigned to the currently measured electrical potential. The temperature
dependency of the calibration standard is considered during the assignment.

Temperatue dependency of Hamilton pH buffer solutions

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

0

10

20

30

40

50

60

Temperature [°C]

p

H

Figure 2.7.3.1.2: Temperature dependency of the pH value of HAMILTON DURACAL calibration

standards.

If the sensor does not find any corresponding calibration standard, the bit representing the
corresponding status

no matching calibration standard

”is set in the calibration status register.

Note that only the nominal value of the calibration standard is used for calibration. The actual value of
the calibration solution, which may deviate from the nominal value, is not taken into account and cannot
be set by the operator.

Standard Calibration with Manual Selection of the Calibration Standard

If the operator knows in what calibration standard the sensor is immersed, he can initiate the calibration
procedure by means of setting the pH value to the actual value of the calibration standard. The sensor
now screens the list of calibration standards that are allowed for manual selection (register 9530). If the
pH value entered by the operator fits in the allowed pH range of one of the allowed calibration
standards, the entered pH value is assigned to the currently measured electrical potential. The
temperature dependency of the calibration standard is considered during the assignment.

If the sensor does not find any corresponding calibration standard, the bit representing the
corresponding status

no matching calibration standard

”is set in the calibration status register.

Note: using manual selection only, the actual pH value of the calibration standard can be set. However,
the actual value must be within the given tolerance of the standard in use.