2.7.4 Reading the Calibration Status

2.7.4.1

Reading the Calibration Status of CP1 and CP2

A standard calibration is not always successful. In order to analyze what has gone wrong, two different
calibration status registers can be read:

 Register 5158 for CP1

 Register 5190 for CP2

Note

Registers 5158 and 5190 contain the same information!

Bit #

Hex value

Definition

0 (LSB)

0x00000001

CP1: difference between CP1 and CP2 < pH 1.0

1

0x00000002

CP1: no matching calibration standard

2

0x00000004

CP1: actual temperature reading is too low

3

0x00000008

CP1: actual temperature reading is too high

4

0x00000010

CP1: temperature reading during calibration is not stable

5

0x00000020

CP1: offset a pH 7 is too low or slope is too low

6

0x00000040

CP1: offset a pH 7 is too high or slope is too high

7

0x00000080

CP1: pH reading during calibration is not stable

8

0x00000100

CP2: difference between CP2 and CP1 < pH 1.0

9

0x00000200

CP2: no matching calibration standard

10

0x00000400

CP2: actual temperature reading is too low

11

0x00000800

CP2: actual temperature reading is too high

12

0x00001000

CP2: temperature reading during calibration is not stable

13

0x00002000

CP2: offset a pH 7 is too low or slope is too low

14

0x00004000

CP2: offset a pH 7 is too high or slope is too high

15

0x00008000

CP2: pH reading during calibration is not stable

not available

24

0x01000000

CP6: out of calibration range

25

0x02000000

CP6: out of range

26

0x04000000

CP6: active

27

0x08000000

CP6: initial measurement

28

0x10000000

CP6: assigned
not available

30

0x40000000

CP2: incorrect measurement unit

31

0x80000000

CP1: incorrect measurement unit

Figure 2.7.4.1.1: Definition of the status for register 5158, 5190 and 5318 (see Figure 2.7.4.1.2 and

2.7.4.2.1).

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus
function
code

Read
access

Write
access

5158

6

Status CP1
(see figure
2.7.4.1.1)

Physical unit of
the last
successful
calibration CP1
(always pH)

pH value of the
last successful
calibration CP1

3, 4

U/A/S

none

5190

6

Status CP2
(see figure
2.7.4.1.1)

Physical unit of
the last
successful
calibration CP2
(always pH)

pH value of the
last successful
calibration CP2

3, 4

U/A/S

none

Figure 2.7.4.1.2: Definition of register 5158 for CP1 and register 5190 for CP2.