Hamilton Electronics EPHUM011 User Manual

Page 49

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 49 / 67

2.7.7 Currently active Calibration Parameters part 3

In register 5520, 5528 and 5560 the pH value of the used calibration standards, the electrical potential
of the pH sensor and the temperature upon calibration are stored.

Start
register

Number
of
registers

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

5520

8

pH value of
calibration
standard at
CP1 [pH]

Electrical
potential at
CP1
[mV]

Electrical
potential at
CP1
[mV]

free

3, 4

A/S

none

5528

8

pH value of
calibration
standard at
CP2 [pH]

Electrical
potential at
CP2
[mV]

Electrical
potential at
CP2
[mV]

free

3, 4

A/S

none

5560

8

pH value of
product at
CP6 [pH]

Electrical
potential at
CP6 [mV]

Electrical
potential at
CP6 [mV]

free

3, 4

A/S

none

Figure 2.7.7.1: Definition of register 5520, 5528 and 5560.

Command: Act calibration CP1

Modbus address:

5520

Length:

8

Type:

3

Read

Parameter:

pH CP1 [pH]

Voltage CP1 [mV]

Temp CP1 [K]

free

Format:

Float

Float

Float

Float

Value:

4.003401

179.927

297.1378

0

Figure 2.7.7.2: Example to read the actual calibration values of CP1.

Command: Act calibration CP2

Modbus address:

5528

Length:

8

Type:

3

Read

Parameter:

pH CP2 [pH]

Voltage CP2 [mV]

Temp CP2 [K]

free

Format:

Float

Float

Float

Float

Value:

7.006804

3.099747

296.6901

0

Figure 2.7.7.3: Example to read the actual calibration values of CP2.

Command: Act calibration CP6

Modbus address:

5560

Length:

8

Type:

3

Read

Parameter:

pH CP6 [pH]

Voltage CP6 [mV]

Temp CP6 [K]

free

Format:

Float

Float

Float

Float

Value:

7.1

5.10469

298.3302

0

Figure 2.7.7.4: Example to read the actual calibration values of CP6.

2.7.8 Currently active Calibration Parameters part 4

For standard calibration (CP1 / CP2) register 5448 documents offset at pH7 and slope:

Start
register

Number
of
registers

Reg1 / Reg2
(Float)

Reg3 / Reg4
(Float)

Reg5 / Reg6
(Float)

Modbus
function
code

Read
access

Write
access

5448

6

Offset at
pH 7 [mV]

Slope (25 °C)
[mV/pH]

Reference
temperature [K]

3, 4

U/A/S

none

Figure 2.7.8.1: Definition of register 5448.

Command: Calculated cali values

Modbus address:

5448

Length:

6

Type:

3

Read

Parameter:

Offset at pH 7
[mV]

Slope
[mV/pH]

Ref temp
[K]

Format:

Float

Float

Float

Value:

3.607782

-59.47631

298.15

Figure 2.7.8.2: Example to read register 5448: offset at pH 7 is 3.6 mV; slope is -59.5 mV/pH; reference

temperature is 298.15 K (=25°C)