pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 50 / 67

2.7.9 Special Commands for Calibration with VISICAL

The VISICAL calibration device allows calibration of pH ARC Sensors at CP1 or CP2. The pH ARC
Sensor

s associated calibration parameters for CP1 and CP2 are those predefined and stored in

corresponding registers of the sensor.
Register 5164 defines the pH value for CP1 and register 5196 defines the pH value for CP2, which are
only valid for use with VISICAL. The same calibration limits for the pH value are used as for standard
calibration at CP1 and CP2 (register 5152 and 5184 respectively).

Attention:

 It is not possible to perform a product calibration using VISICAL.

 Physical unit is fixed to pH by definition.

Start
register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

5180

2

pH value at CP1 (default: pH 4.00)

3, 4, 16

U/A/S

S

5212

2

pH value at CP2 (default: pH 7.00)

3, 4, 16

U/A/S

S

Figure 2.7.9.1: Definition of register 5180 for CP1 and 5212 for CP2.

Command: VISICAL CP1

Modbus address:

5180

Length:

2

Type:

3

Read

Parameter:

Value [pH]

Format:

Float

Value:

4.00

Figure 2.7.9.2: Example to read the pH value valid for CP1. It is 4.00. Accordingly, the next time when a

calibration is started using VISICAL at LOW, a calibration with manual selection is performed,
using the calibration standard at pH 4. The operator has to make sure that - within the selected
set of calibration standards - a standard at pH 4 is selected for manual selection.

Command: VISICAL CP2

Modbus address:

5212

Length:

2

Type:

16

Write

Parameter:

Value [pH]

Format:

Float

Value:

7.00

Figure 2.7.9.3: Example to set the pH value valid for CP2.

Command: VISICAL CP2

Modbus address:

5212

Length:

2

Type:

3

Read

Parameter:

Value [pH]

Format:

Float

Value:

7.00

Figure 2.7.9.4: Example to read the pH value valid for CP2. It is 7.00. Accordingly, the next time when a

calibration is started using VISICAL at HIGH, a calibration with manual selection is performed,
using the calibration standard at pH 7. The operator has to make sure that - within the selected
set of calibration standards - a standard at pH 7 is selected for manual selection.