Hamilton Electronics EPHUM011 User Manual

Page 52

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 52 / 67

2.7.10.2 Definitions for Individual Sets of Calibration Standards

Once the operator has selected a set of calibration standards (register 9474) the register 9504 and
following give all information on the selected set of calibration standards.

Start
register

Number of
registers

Reg1

… Reg8

(16 ASCII characters)

Modbus
function code

Read
access

Write
access

9504

8

Manufacturer of the selected set

3, 4

U/A/S

none

9512

8

Info 1 of the selected set

3, 4

U/A/S

none

9520

8

Info 2 of the selected set

3, 4

U/A/S

none

Figure 2.7.10.2.1: Definition of registers 9504 to 9520.

Command: manufacturer of set

Modbus address:

9504

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

HAMILTON

Figure 2.7.10.2.2: Example to read the description (in this case, the set 0x01 is selected)

Within one set of calibration standards, a maximum of 12 calibration standards are available. In the
following registers, the details of each calibration standard is given:

Start register

Description

9536

Nominal value and tolerance of calibration standard 1

9552

Nominal value and tolerance of calibration standard 2

9568

Nominal value and tolerance of calibration standard 3

9584

Nominal value and tolerance of calibration standard 4

9600

Nominal value and tolerance of calibration standard 5

9616

Nominal value and tolerance of calibration standard 6

9632

Nominal value and tolerance of calibration standard 7

9648

Nominal value and tolerance of calibration standard 8

9664

Nominal value and tolerance of calibration standard 9

9680

Nominal value and tolerance of calibration standard 10

9696

Nominal value and tolerance of calibration standard 11

9712

Nominal value and tolerance of calibration standard 12

Figure 2.7.10.2.3: Definition for the register range from 9536 until 9720.

Start
register

Number
of
registers

Reg1 /
Reg2
(Float)

Reg3 /
Reg4
(Float)

Reg5 /
Reg6
(Float)

Reg7 /
Reg8
(Float)

Modbus
function
code

Read
access

Write
access

9536,
9552,

8

Calibr.
standard
nominal
value
[pH]

Tolerance
of nominal
value (±)
[pH]

Nominal
electrical
potential
[mV]

Tolerance
of electrical
potential
(±) [mV]

3, 4

U/A/S

none

Figure 2.7.10.2.4: Definition for registers 9536 until 9720.

Command: Read standard 4

Modbus address:

9584

Length:

8

Type:

3

Read

Parameter:

Nominal value
[pH]

Tolerance
(±) [pH]

Nominal electrical
potential
[mV]

Tolerance of
electrical potential
(±) [mV]

Format:

Float

Float

Float

Float

Value:

4.01

0.02

180

80

Figure 2.7.10.2.5: Example to read the values for the calibration standard 4.