What happens to the pH ARC Sensor

s calibration function upon product calibration (CP6)?

A product calibration adds an offset to the linear calibration function defined by the standard calibration
at CP1 and CP2.

Product calibration of ARC pH sensors

-200

-150

-100

-50

0

50

100

150

200

5

6

7

8

9

10

pH

E

le

c

tr

ic

a

lp

o

te

n

ti

a

l[

m

V

]

Standard Calibration function

Product Calibration function

CP1

CP2

CP6

Figure 2.7.3.2.1: Effect of the product calibration CP6 on an existing standard calibration function

defined by CP1 and CP2.

The operator starts with a Standard Calibration with calibration points CP1 and CP2:
CP1: pH value of calibration standard: 6

electrical potential: 64.2 mV

CP2: pH value of calibration standard: 9.21

electrical potential: -125.2 mV

The sensor internally calculates the calibration function, using the calibration points

CP1 and

CP2. The resulting calibration function, compensated to the standard temperature 25°C, is
shown as a straight line. The calibration function is described by two parameters: the offset at pH
7 and the slope.
Some weeks later, the operator believes that the Standard Calibration function is not correct
anymore. As the process is running and he is not able to perform a standard calibration under
defined conditions in the lab, he decides to perform a product calibration CP6, in other words
adjusting the standard calibration function to the process conditions:

CP6: pH value of product: 7.2

electrial potential: 5 mV

The sensor internally adds an offset to the calibration curve. The slope remains unchanged.

Another special feature of this calibration point is to switch off and back on again a product calibration.
These functions are called

restore standard calibration

”and “

restore product calibration

.

Note:
The sensor

s internal criteria for a successful product calibration are:

 the sensor is currently in an environment corresponding to the pH ARC Sensors measurement

range.

 the manually assigned pH value does not deviate more than 2 pH units from the value

measured prior the product calibration