2.7.5 Currently active Calibration Parameters part 1

In registers 5164 (CP1), 5196 (CP2) and 5324 (CP6) the currently active calibration parameters part 1
are stored. These registers contain the values for temperature, number of calibrations and operating
hour upon calibration.

Start
register

Number
of
registers

Reg1 / Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

5164

8

Unit of
temperature for
CP1 (bitwise
defined)

Value of
temperature
of CP1

Number of
calibrations
at CP1

Operating
hour for
CP1

3, 4

U/A/S

none

5196

8

Unit of
temperature for
CP2 (bitwise
defined)

Value of
temperature
of CP2

Number of
calibrations
at CP2

Operating
hour for
CP2

3, 4

U/A/S

none

5324

8

Unit of
temperature for
CP6 (bitwise
defined)

Value of
temperature
of CP6

Number of
calibrations
at CP6

Operating
hour for
CP6

3, 4

U/A/S

none

Figure 2.7.5.1: Definition of register 5164 for CP1, 5196 for CP2 and 5324 for CP6.

Command: Calibration CP1 values

Modbus address:

5164

Length:

8

Type:

3

Read

Parameter:

Unit of temperature

Temperature

Number of cali

Operating hour

Format:

Hex

Float

Decimal

Float

Value:

0x00000004

24.35184

6

23.78

Figure 2.7.5.2: Example to read the calibration values for CP1. The physical unit is °C (0x00000004),

the temperature is 24.35184 °C, the number of calibrations at CP1 is 6 and the operating hour is
23.78 h.

Command: Calibration CP2 values

Modbus address:

5196

Length:

8

Type:

3

Read

Parameter:

Unit of temperature

Temperature

Number of cali

Operating hour

Format:

Hex

Float

Decimal

Float

Value:

0x00000004

24.37691

5

16.45

Figure 2.7.5.3: Example to read the calibration values for CP2. The physical unit is °C (0x00000004),

the temperature is 24.37691 °C, the number of calibrations at CP1 is 5 and the operating hour is
16.45 h.

Command: Calibration CP6 values

Modbus address:

5324

Length:

8

Type:

3

Read

Parameter:

Unit of temperature

Temperature

Number of cali

Operating hour

Format:

Hex

Float

Decimal

Float

Value:

0x00000004

29.93368

12

379.5167

Figure 2.7.5.4: Example to read the calibration values 1 for CP6. The physical unit is °C (4), the

temperature is 29.93 (°C), the number of calibrations at CP1 is 12 and the operating hour is
379.51 (h).

2.7.6 Currently active Calibration Parameters part 2

Registers 5172 (CP1), 5204 (CP2) and 5332 (CP6) are not defined for pH ARC Sensors, as they
document atmospheric pressure and salinity used for VISIFERM DO / VISIFERM DO ARC Sensors
only.