pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 47 / 67

The here performed assignment was successful. The sensor is running using a valid product
calibration.

2.7.4.2.3 Product calibration: Cancel

Calibration status after cancelling an active product calibration:

Command: Calibration status CP6

Modbus address:

5318

Length:

6

Type:

3

Read

Parameter:

Status

Unit

Value

Format:

Hex

Hex

Float

Value:

0x00000000

0x00001000

4.5

Figure 2.7.4.2.3.1: Example to read the calibration status of CP6 after having performed a cancel

command at CP6.
The status reports no messages. The last successful calibration at CP6 has been performed at
pH 4.5.
The sensor is running on a valid standard calibration and no product calibration is stored.

2.7.4.2.4 Product calibration: Restore standard calibration

Calibration status after restoring a standard calibration from an active product calibration:

Command: Calibration status CP6

Modbus address:

5318

Length:

6

Type:

3

Read

Parameter:

Status

Unit

Value

Format:

Hex

Hex

Float

Value:

0x10000000

0x00001000

4.5

Figure 2.7.4.2.4.1: Example to read the calibration status of CP6 after having restored the standard

calibration from an active product calibration (CP6).
The status says:

CP6 assigned

”(0x10000000). The last successful calibration at CP6 has been

performed at pH 4.5.
The sensor is running on a valid standard calibration but a valid product calibration is still
available in the sensor.

2.7.4.2.5 Product calibration: Restore product calibration

Calibration status after restoring an available product calibration from an active standard calibration:

Command: Calibration status CP6

Modbus address:

5318

Length:

6

Type:

3

Read

Parameter:

Status

Unit

Value

Format:

Hex

Hex

Float

Value:

0x14000000

0x00001000

4.5

Figure 2.7.4.2.5.1: Example to read the calibration status of CP6 after having restored an available

product calibration (CP6) from an active standard calibration (CP1 / CP2).
The status says:

CP6: active

”(0x04000000) and “

CP6: assigned

”(0x10000000). The last

successful calibration corresponding to the here performed assignment has been performed at
pH 4.5.
The sensor is running on a valid product calibration again.