2.6.3 PA12: Moving Average R

pH ARC Sensors allow to have a separate moving average on secondary measurement values:

 Glass resistance

 Reference resistance

The moving average can be applied on 1 to 16 3-s measurement values. The default value is 4.
Especially if high resistances are measured, it is recommended to choose a higher moving average.

2.6.3.1

Description of PA12 (Moving Average R)

In register 3456, a plain text ASCII description of PA12 is given.

Start
register

Number of
registers

Reg1

… Reg8

16 ASCII characters

Modbus
function code

Read
access

Write
access

3456

8

Description of PA12

3, 4

U/A/S

none

Figure 2.6.3.1.1: Definition of register 3456.

Command: Moving average text

Modbus address:

3456

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

Moving average R

Figure 2.6.3.1.2: Example to read the description for

Moving average R

.

2.6.3.2

Selecting the Physical Unit and Writing the Value for PA12

In register 3464, the available physical units for PA12 are defined.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

3464

2

Available physical units for PA12

3, 4

U/A/S

none

Figure 2.6.3.2.1: Definition of register 3368.

Command: Moving average av. units

Modbus address:

3464

Length:

2

Type:

3

Read

Parameter:

Units

Format:

Hex

Value:

0x01

Figure 2.6.3.2.2: Example to read the available physical units for PA12. The only one available here is

none

”(0x01). For the definition of the physical units see chapter 2.5.1.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

3466

4

Select physical unit
for PA12

Value for PA12
(1-16, default: 4)

16

none

S

Figure 2.6.3.2.3: Definition of register 3466. Only one bit for the physical unit can be set. PA12 can be

set to the values 1-16.