pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 18 / 67

2.5

Measurement

2.5.1 Definition of Measurement Channels and Physical Units

The pH ARC Sensor Modbus register structure allows the definition of 6 individual Primary
Measurement Channels (PMC), and 16 individual Secondary Measurement Channels (SMC).

Bit #

Hex code

Description

Definition

0 (LSB)

0x000001

PMC1

pH

1

0x000002

PMC2

not available
not available

4

0x000010

PMC5

not available

5

0x000020

PMC6

Temperature

6

0x000040

SMC1

R glass

7

0x000080

SMC2

R reference

8

0x000100

SMC3

R auxiliary

9

0x000200

SMC4

E pH vs. ref

10

0x000400

SMC5

E SG vs. ref

11

0x000800

SMC6

E aux vs. ref

12

0x001000

SMC7

E reference

13

0x002000

SMC8

pH act

14

0x004000

SMC9

T act

15

0x008000

SMC10

not available

21 (MSB)

0x200000

SMC16

not available

Figure 2.5.1.1: full list of PMC1 to 6 and SMC1 to 16.

In Register 2048, the available PMC and SMC are defined for a specific pH ARC Sensors and a
specific operator level.

Start
register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

2048

2

Available measurement channels
PMC and SMC (bitwise set)

3, 4

U/A/S

none

Figure 2.5.1.2: Definition of register 2048.

Command: Avail. PMC and SMC

Modbus address:

2048

Length:

2

Type:

3

Read

Parameter:

Avail. PMC and
SMC

Format:

Hex

Value:

0x06E1

Figure 2.5.1.3: Example to read Reg. 2048 for Polilyte Plus ARC.

In case of operator A/U, the value 0x06E1 is returned. In other words the following PMC and
SMC are available to A/U: PMC1 / PMC6 / SMC1 / SMC2 / SMC4 / SMC5
In case of operator S, the value 0x066E1 is returned. In other words the following PMC and SMC
are available to S:

PMC1 / PMC6 / SMC1 / SMC2 / SMC4 / SMC5 / SMC8 / SMC9