3abbreviations, 10 system commands – Hamilton Electronics EPHUM011 User Manual

Page 66

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 66 / 67

Command: Info user

Modbus address:

1568

Length:

8

Type:

16

Write

Parameter:

Text

Format:

Character

Value:

Hello World

Figure 2.9.3.2: Example to write 16 ASCII characters to register 1568 with operator A or S.

Command: Info user

Modbus address:

1568

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

Hello World

Figure 2.9.3.3: Example to read the register 1568 (written in figure 2.9.1.6).

2.10 System Commands

2.10.1 Recall Sensor’

s Factory Settings

Using register 8192 you can recall the sensor manufacturer values (interfaces, calibration data and
passwords), except the SIP and CIP data which remain unchanged. By sending the recall value

732255

, all configuration values will be set to default.

Start
register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

8192

2

Recall by value 732255

16

none

S

Figure 2.10.1.1: Definition of register 8192.

Command: Recall

Modbus address:

8192

Length:

2

Type:

16

Write

Parameter:

Recall

Format:

Decimal

Value:

732255

Figure 2.10.1.2: Example to write the restore command.

3

Abbreviations

AO

Analog Output Interface

CP

Calibration Point

ECS

Electrochemical Sensor Interface

PMC

Primary Measurement Channel

SMC

Secondary Measurement Channel