8 sensor status – Hamilton Electronics EPHUM011 User Manual

Page 57

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 57 / 67

2.8

Sensor Status

2.8.1 Temperature Ranges

In registers 4608, 4612 and 4616 three different temperature ranges are defined:

 Operation –in this range the sensor will work properly (current output, Modbus

communication), except the measurement, which is stopped until the temperature is back in
the measurement range. In this case the last value of measurement will be frozen and sent to
analog interfaces.

 Measurement –in this range the sensor is able to measure.

 Calibration –in this range the sensor can be calibrated.

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Modbus
function
code

Read
access

Write
access

4608

4

Operating
temperature min [°C]

Operating
temperature max [°C]

3, 4

U/A/S

none

4612

4

Measurement
temperature min [°C]

Measurement
temperature max [°C]

3, 4

U/A/S

none

4616

4

Calibration
temperature min [°C]

Calibration
temperature max [°C]

3, 4

U/A/S

none

Figure 2.8.1.1: Definition of register 4608, 4612 and 4616.

Command: Operating T range

Modbus address:

4608

Length:

4

Type:

3

Read

Parameter:

Operating T min
[°C]

Operating T max
[°C]

Format:

Float

Float

Value:

-20

130

Figure 2.8.1.2: Example to read the operating temperature values min and max.

Command: Measurement T range

Modbus address:

4612

Length:

4

Type:

3

Read

Parameter:

Measurement T
min [°C]

Measurement T
max [°C]

Format:

Float

Float

Value:

-20

130

Figure 2.8.1.3: Example to read the measurement temperature values min and max.

Command: Calibration T range

Modbus address:

4616

Length:

4

Type:

3

Read

Parameter:

Calibration T min
[°C]

Calibration T max
[°C]

Format:

Float

Float

Value:

5

50

Figure 2.8.1.4: Example to read the calibration temperature values min and max.