Hamilton Electronics EPHUM011 User Manual

Page 29

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 29 / 67

2.6.2.3

Reading all Values for PA9

By reading register 3370, the active physical unit of measurement, the selected value, and the min and
max values can be read.

Start
register

Number of
registers

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

3370

8

Physical
unit

Current
value

Min
value

Max
value

3, 4

U/A/S

none

Figure 2.6.2.3.1: Definition of register 3370.

Command: Moving average

Modbus address:

3370

Length:

8

Type:

3

Read

Parameter:

Unit

Value

Min value

Max value

Format:

Hex

Decimal

Decimal

Decimal

Value:

0x01

10

1

16

Figure 2.6.2.3.2: Example to read PA9. The physical unit is 0x01 (

none

), the value is 10 and the limit

is 1 to 16.