624300/01

page 32 / 67

2.7

Calibration

2.7.1 Available Calibration Points

In register 5120, the available number of Calibration Points (CP) for Primary Measurement Channel 1
(PMC1) is defined. 8 individual CP are theoretically possible.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

5120

2

Available number of CP for PMC1
(see figure 2.7.1.2)

3, 4

U/A/S

none

Figure 2.7.1.1: Definition of register 5120.

Bit #

Hex value

Description

Definition in ARC Sensors

0 (LSB)

0x01

CP1

Calibration Point 1

1

0x02

CP2

Calibration Point 2

2

0x04

CP3

not available

not available

5

0x20

CP6

Product Calibration

6

0x40

CP7

not available

7 (MSB)

0x80

CP8

not available

Figure 2.7.1.2: Bitwise definition of CP1 to CP8.

Command: Available cali points

Modbus address:

5120

Length:

2

Type:

3

Read

Parameter:

Points

Format:

Hex

Value:

0x23

Figure 2.7.1.3: Example to read the available CPs. 0x23 = 0x01 (CP1) + 0x02 (CP2) + 0x20 (CP6).

Standard Calibration at 2 calibration points

-200

-150

-100

-50

0

50

100

150

200

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

pH value

e

le

c

tr

ic

a

lp

o

te

n

ti

a

lg

la

s

s

v

s

.r

e

f

[m

V

]

CP2

CP1

Figure 2.7.1.4: pH ARC Sensors allow 3 calibration points:

CP1 and CP2 are used for standard calibration (shown in this figure).
The product calibration CP6 is used to adjust the standard calibration function to specific process
conditions (the effect of CP6 is shown in Figure 2.7.3.2.1).