pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 28 / 67

2.6.2.1

Description of PA9 (Moving Average)

In register 3360, a plain text ASCII description of PA9 is given.

Start
register

Number of
registers

Reg1

… Reg8

16 ASCII characters

Modbus
function code

Read
access

Write
access

3360

8

Description of PA9

3, 4

U/A/S

none

Figure 2.6.2.1.1: Definition of register 3360.

Command: Moving average text

Modbus address:

3360

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

Moving average

Figure 2.6.2.1.2: Example to read the description for

Moving average

.

2.6.2.2

Selecting the Physical Unit and Writing the Value for PA9

In register 3368, the available physical units for PA9 are defined.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

3368

2

Available physical units for PA9

3, 4

U/A/S

none

Figure 2.6.2.2.1: Definition of register 3368.

Command: Moving average av. units

Modbus address:

3368

Length:

2

Type:

3

Read

Parameter:

Units

Format:

Hex

Value:

0x01

Figure 2.6.2.2.2: Example to read the available physical units for PA9. The only one available here is

none

”(0x01). For the definition of the physical units see chapter 2.5.1.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

3370

4

Select physical unit
for PA9

Value for PA9
(1-16, default: 2)

16

none

S

Figure 2.6.2.2.3: Definition of register 3370. Only one bit for the physical unit can be set. PA9 can be

set to the value 1-16. A value of 1 does not influence the response time of the sensor, a value of
16 increases the response time of the sensor to 48 s.

By writing to register 3370 the active physical unit for PA9 can be selected by choosing one of the
physical units that are defined in register 3368. The value of the parameter can be set as well.

Command: Moving average

Modbus address:

3370

Length:

4

Type:

16

Write

Parameter:

Unit

Value

Format:

Hex

Decimal

Value:

0x01

12

Figure 2.6.2.2.4: Example to set the physical unit of PA9 to

none

”(0x01) and the value of the moving

average to 12.