Hamilton Electronics EPHUM011 User Manual

Page 42

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 42 / 67

2.7.3.2.2 Product calibration: Assignment

After successful initial measurement a correct value must be assigned to the initially stored
measurement data.
This is achieved by writing the correct calibration value to register 5322.

Start
register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

5322

2

pH value [pH]

16

none

A/S

Figure 2.7.3.2.2.1: Definition of register 5322

Command: CP6: Assignment

Modbus address:

5322

Length:

2

Type:

16

Write

Parameter:

Value

Format:

Float

Value:

7.2

Figure 2.7.3.2.2.2: Example to assign a calibration value to the above performed initial measurement.

This is achieved by writing the correct pH value.

From now on the sensor is measuring using the here performed product calibration.

The calibration status (register 5318) is 0x14000000 meaning that a correct value has been assigned
and that the product calibration is active (see figure 2.7.4.1.1).

2.7.3.2.3 Product calibration: Cancel

To cancel an active product calibration or an active initial measurement the command 0x02 is written to
register 5340.

Command: CP6: Cancel

Modbus address:

5340

Length:

2

Type:

16

Write

Parameter:

Command

Format:

Hex

Value:

0x02

Figure 2.7.3.2.3.1: Example to cancel an active product calibration or an initial measurement. Writing

the command 0x02 (cancel) to register 5340.

Performing this action the product calibration or any initial measurements are canceled. The values of
the prior product calibration are removed from the sensor

s memory. From now on the sensor is

measuring using its prior CP1 / CP2 standard calibration.

The sensor

s calibration status (register 5318) will be reading 0x00 again (see figure 2.7.4.1.1).