pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 17 / 67

2.4.6 Reading the Internally Measured Output Current

Reg. 4414 / 4542 provides internal parameters of AO1 / AO2:

 the setpoint to which the current is regulated in a closed loop control

 the electrical current the sensor is measuring to feed the closed loop control

These values are helpful in order to compare against the externally measured electrical current.

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

4414

4

Set point
[mA]
AO1

Internally measured
[mA]
AO1

3, 4

U/A/S

none

4542

4

Set point
[mA]
AO2

Internally measured
[mA]
AO2

3, 4

U/A/S

none

Figure 2.4.6.1: Definition of register 4414 / 4542.

Command: Internal values AO1

Modbus address:

4414

Length:

4

Type:

3

Read

Parameter:

Set point
[mA]

Internally measured
[mA]

Format:

Float

Float

Value:

9.99186

9.99742

Figure 2.4.6.2: Example to read the internal values of AO1, depending on the analog interface mode.