Hamilton Electronics EPHUM011 User Manual

Page 51

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 51 / 67

2.7.10 Calibration Standards

2.7.10.1 Available Sets of Calibration Standards

pH ARC Sensors can store and operate six different sets of calibration standards each having max 12
calibration standards. Before calibration the operator has to select one defined set to operate with.
Default setting is the HAMILTON set of calibration standards.
The operator can change the selection of sets at any time, without compromising prior calibration data
at CP1 / CP2.

In register 9472 the available sets of calibration standards are defined.

Start
register

Number of
registers

Reg1 / Reg2
(Bit, see Figure 2.7.10.1.2)

Modbus
function code

Read
access

Write
access

9472

2

Available calibration standard sets

3, 4

U/A/S

none

Figure 2.7.10.1.1: Definition of register 9472.

Bit #

Hex value

Description

Definition

0 (LSB)

0x00000001

Set 1

HAMILTON

1

0x00000002

Set 2

MERCK TITRISOL

2

0x00000004

Set 3

DIN 19267

3

0x00000008

Set 4

NIST Standard

4

0x00000010

Set 5

METTLER-TOLEDO

5

0x00000020

Set 6

RADIOMETER

Figure 2.7.10.1.2: Definition of available sets of calibration standards.

Command: Available cali sets

Modbus address:

9472

Length:

2

Type:

3

Read

Parameter:

Calibration sets

Format:

Hex

Value:

0x0000003F

Figure 2.7.10.1.3: Example to read the available calibration sets: Set 1 (0x00000001) + set 2

(0x00000002) + set 3 (0x00000004) + set 4 (0x00000008) + set 5 (0x00000010) + set 6
(0x00000020), in total 0x0000003F.

In register 9474 the selected set of calibration standards is defined.

Start
register

Number of
registers

Reg1 / Reg2
(Bit, see Figure 2.7.10.1.2)

Modbus
function code

Read
access

Write
access

9474

2

Selected set of calibration standard

3, 4

U/A/S

S

Figure 2.7.10.1.4: Definition of register 9474. Only one bit can be set.

Command: Selected set of cal stand.

Modbus address:

9474

Length:

2

Type:

3

Read

Parameter:

Calibration set

Format:

Hex

Value:

0x00000001

Figure 2.7.10.1.5: Example to read the selected set of calibration standards. Set 1 (HAMILTON)

(0x00000001) is active.

Command: Select set of cal stand.

Modbus address:

9474

Length:

2

Type:

16

Write

Parameter:

Calibration set

Format:

Hex

Value:

0x04

Figure 2.7.10.1.6: Example to set the calibration standard set to DIN 19267 (0x04).