pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 20 / 67

2.5.2 Primary Measurement Channel 1 (pH)

2.5.2.1

Description of PMC1

In register 2080, a plain text ASCII description of PMC1 is given.

Start
register

Number of
registers

Reg1

… Reg8

16 ASCII characters

Modbus
function code

Read
access

Write
access

2080

8

Description of PMC1

3, 4

U/A/S

none

Figure 2.5.2.1.1: Definition of register 2080.

Command: PMC 1 text

Modbus address:

2080

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

pH

Figure 2.5.2.1.2: Example to read the description. It is

pH

.

2.5.2.2

Selecting the Physical Unit for PMC1

In register 2088, the available physical units for this channel are defined.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

2088

2

Available physical units of PMC1

3, 4

U/A/S

none

Figure 2.5.2.2.1: Definition of register 2088.

Command: PMC1 available units

Modbus address:

2088

Length:

2

Type:

3

Read

Parameter:

Units

Format:

Hex

Value:

0x201000

Figure 2.5.2.2.2: Example to read the available physical units of PMC1: pH (0x001000) + mV

(0x200000), total 0x201000.

In register 2090, the active physical unit for this channel can be selected, by choosing one of the
physical units that are defined in register 2088.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

2090

2

Selected active physical unit for the
PMC1

16

none

S

Figure 2.5.2.2.3: Definition of register 2090. Only one bit can be set.

Command: PMC1 set unit

Modbus address:

2090

Length:

2

Type:

16

Write

Parameter:

Unit

Format:

Hex

Value:

0x1000

Figure 2.5.2.2.4: Example to set the physical unit of PMC1 to pH (0x1000).

Attention:

Changing the physical unit has also an influence on the output of AO1 / AO2, as the same physical unit
is active for the analog outputs. All limits of the 4-20 mA analog output have to be redefined after
changing the physical unit!