pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 31 / 67

By writing to register 3466, the active physical unit for PA12 can be selected, by choosing one of the
physical units that are defined in register 3464. The value of the parameter can be set as well.

Command: Moving average

Modbus address:

3466

Length:

4

Type:

16

Write

Parameter:

Unit

Value

Format:

Hex

Decimal

Value:

0x01

7

Figure 2.6.3.2.4: Example to set the physical unit of PA12 to

none

”(0x01) and the value of the moving

average R to 7.

2.6.3.3

Reading all Values for PA12

By reading register 3466, the active physical unit of measurement, the selected value, and the min and
max values can be read.

Start
register

Number of
registers

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

3466

8

Physical
unit

Current
value

Min
value

Max
value

3, 4

U/A/S

none

Figure 2.6.3.3.1: Definition of register 3466.

Command: Moving average

Modbus address:

3466

Length:

8

Type:

3

Read

Parameter:

Unit

Value

Min value

Max value

Format:

Hex

Decimal

Decimal

Decimal

Value:

0x01

7

1

16

Figure 2.6.3.3.2: Example to read PA12. The physical unit is 0x01 (

none

), the value is 7, and the

limits are 1 to 16.