2ph arc sensor commands in modbus rtu, 1 general, 2 operator levels and passwords – Hamilton Electronics EPHUM011 User Manual

Page 5

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 5 / 67

2

pH ARC Sensor Commands in Modbus RTU

2.1

General

In order to communicate with a pH ARC Sensor over Modbus RTU protocol a Modbus master terminal
application software is needed. The Modbus RTU is an open standard and a number of free and
commercial application toolkits are available.

This manual contains examples and illustrations from WinTECH Modbus Master ActiveX Control tool:
WinTECH (

www.win-tech.com

)

Modbus Master OCX for Visual Basic

. The Modbus Organisation

(

www.modbus.org/tech.php

) provides other links to a wide variety of Modbus terminal softwares.

In the present manual the addressing of the Modbus registers starts at 1. But the Modbus master
protocol operates with register addresses starting at 0. Usually, the Modbus master software translates
the addressing. Thus, the register address of 2090 will be translated by the Modbus master software to
2089 which is sent to the sensor (Modbus slave).

Attention:

When configuring and calibrating the sensor, please limit write operations to a reasonable number.
More than 100

000 write operations will physically damage the memory of the sensor.

Furthermore, for the Free User Memory Space (see chapter 2.9.3), the write operations are limited to
10

000.

2.2

Operator levels and Passwords

2.2.1 Reading / Setting Operator Level

A pH ARC Sensor can be operated in three different operator levels. Each operator level allows a
defined access to a specific set of commands.

Abbreviation

Description

Code (hex)

Password (decimal)

U

User (lowest level)

0x03

0

A

Administrator

0x0C

18111978

S

Specialist

0x30

16021966

Figure 2.2.1.1: Definition of operator level and default passwords

At each power up or processor reset, the operator level falls back to the default level U.

The active operator level can be read and written in register 4288.

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

4288

4

Operator Level

Password

3, 4, 16

U/A/S

U/A/S

Figure 2.2.1.2: Definition of register 4288.

Command: Active operator level

Modbus address:

4288

Length:

4

Type:

3

Read

Parameter:

Operator level

Password

Format:

hex

decimal

Value:

0x03

0

Figure 2.2.1.3: Example to read the active operator level (function code 3, start register address 4288,

number of registers 4): The active operator level is 0x03 (User). The sensor does not report the
password. The value 0 is returned instead.