Hamilton Electronics EPHUM011 User Manual

Page 13

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 13 / 67

2.4.5.5

Reading the Selected Physical Unit for Analog Interface

Start
register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

4376

2

Selected physical unit of AO1
(see chapter 2.5.1)

3, 4

U/A/S

none

4504

2

Selected physical unit of AO2
(see chapter 2.5.1)

3, 4

U/A/S

none

Figure 2.4.5.5.1: Definition of register 4376 / 4504.

Command: Avail unit AO1

Modbus address:

4376

Length:

2

Type:

3

Read

Parameter:

Available unit

Format:

Hex

Value:

0x001000

Figure 2.4.5.5.2: Example to read the selected unit of the selected PMC of AO1. The value returned is

0x001000

, accordingly, the unit is pH. The physical unit for PMC is defined in Reg. 2090 or

2410 and applies automatically for 4-20 mA output.

2.4.5.6

Defining the Measurement Values for 4, 12 and 20 mA Output

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus
function
code

Read
access

Write
access

4378

6

Measurement
value at
Min Output
Current (4 mA)
for AO1

Measurement
value at
Max Output
Current (20
mA) for AO1

Measurement
value at
Mid Output
Current (12
mA) for AO1

3, 4, 16

U/A/S

S

4506

6

Measurement
value at
Min Output
Current (4 mA)
for AO2

Measurement
value at
Max Output
Current (20
mA) for AO2

Measurement
value at
Mid Output
Current (12
mA) for AO2

3, 4, 16

U/A/S

S

Figure 2.4.5.6.1: Definition of register 4378 / 4506.

Command: MinMaxMid value AO1

Modbus address:

4378

Length:

6

Type:

16

Write

Parameter:

Min value

Max value

Mid value

Format:

Float

Float

Float

Value:

3

10

7

Figure 2.4.5.6.2: Example to set the min value to 3 (for 4 mA), the max value to 10 (for 20 mA) and the

mid value to 7 (for 12 mA). The corresponding physical unit can be read in register 4376 / 4504
and in 2090 / 2410.

Note:
Mid current must always be defined. However, in linear output mode, the mid current value has no
physical meaning and will not affect the 4-20 mA output.