2.5.3.3

Reading the measurement value of PMC6

Register 2410 is also used to read the measurement values of PMC6.

Start
reg.

Num-
ber of
reg.

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Reg9 /
Reg10

Modbus
function
code

Read
access

Write
access

2410

10

Selected
physical
unit

Measure
-ment
value of
PMC6

(1)

Measure
-ment
status

(2)

Min
allowed
value

(1)

Max
allowed
value

(1)

3, 4

U/A/S

none

Figure 2.5.3.3.1: Definition of register 2410. Measurement value of PMC6.

(1)

Value is always in the physical unit defined in register 2410.

(2)

For definition of the status see chapter 2.5.4. All bits set to zero means: no problem.

Command: PMC6 read

Modbus address:

2410

Length:

10

Type:

3

Read

Parameter:

Unit

Value

Status

Min limit

Max limit

Format:

Hex

Float

Hex

Float

Float

Value:

0x04

24.35834

0x00

-20

130

Figure 2.5.3.3.2: Example to read register 2410. Physical unit is set to °C (0x04), PMC6 is 24.35834

°C, Status is 0x00, Min allowed value is -20 °C, Max allowed value is 130 °C.

For definition of the measurement status see chapter 2.5.4.

Attention:

You cannot read selectively the registers 3 and 4 for the measurement value only. You have to read the
entire length of the command (10 registers) and extract the desired information.

2.5.3.4

Input of an Externally Measured Temperature

Unlike to the VISIFERM DO, this feature is not available for pH ARC Sensors.

2.5.4 Definition of the Measurement Status for PMC1 / PMC6

This is the definition of the status registers read in registers 2090 (PMC1) and 2410 (PMC6):

Bit #

Hex code

Description

0 (LSB)

0x01

Temperature out of measurement range (see chapter 2.8.1)

1

0x02

Temperature out of operating range (see chapter 2.8.1)

2

0x04

Calibration status not zero (see chapter 2.7.4)

3

0x08

Warning not zero (see chapter 2.8.3)

4

0x10

Error not zero (see chapter 2.8.4)

Figure 2.5.4.1: Definition of measurement status for Primary Measurement Channels.