pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 65 / 67

Command: Serial number

Modbus address:

1312

Length:

8

Type:

3

Read

Parameter:

Text

Format:

Character

Value:

0001001

Figure 2.9.2.2: Example to read register 1312.

2.9.3 Free User Memory Space

These registers can be used to store any customer specific information in the sensor. There are
different registers which can be read by everybody, but only specific operators can write them.

Start
register

Number
of
registers

Reg1

… Reg8

(16 ASCII characters)

Example of content

Modbus
function
code

Read
access

Write
access

1536

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1544

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1552

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1560

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1568

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1576

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1584

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1592

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1600

8

Measuring point

242111-0001001

3, 4, 16

U/A/S

S

1608

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1616

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1624

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1632

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1640

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1648

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1656

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1664

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1672

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1680

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1688

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1696

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1704

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1712

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1720

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1728

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1736

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1744

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1752

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

Figure 2.9.3.1: Definition of registers containing user information.

An important register is 1600, as it is the description of the measuring point. The information of
this register is displayed on the ARC View Handheld in order to identify individual sensors.

Attention:

The Free User Memory Space is located in a memory which allows in total max 10

000 write

operations.