pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 21 / 67

2.5.2.3

Reading the measurement value of PMC1

Register 2090 is also used to read the measurement values of PMC1.

Start
reg.

Num-
ber of
reg.

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Reg9 /
Reg10

Modbus
function
code

Read
access

Write
access

2090

10

Selected
physical
unit

Measure
-ment
value of
PMC1

(1)

Measure
-ment
status

(2)

Min
allowed
value

(1)

Max
allowed
value

(1)

3, 4

U/A/S

none

Figure 2.5.2.3.1: Definition of register 2090. Measurement value of PMC1.

(1)

Value is always in the physical unit defined in register 2090.

(2)

Definition of the status see chapter 2.5.4. All bits set to zero means: no problem.

Command: PMC1 read

Modbus address:

2090

Length:

10

Type:

3

Read

Parameter:

Unit

Value

Status

Min limit

Max limit

Format:

Hex

Float

Hex

Float

Float

Value:

0x1000

4.02503

0x00

0

14

Figure 2.5.2.3.2: Example to read register 2090. Physical unit is set to pH (0x1000), PMC1 is pH

4.02503, Status is 0x00, Min allowed value is pH 0, Max allowed value is pH 14.

Command: PMC1 read

Modbus address:

2090

Length:

10

Type:

3

Read

Parameter:

Unit

Value

Status

Min limit

Max limit

Format:

Hex

Float

Hex

Float

Float

Value:

0x200000

175.9922

0x00

0

954.6541

Figure 2.5.2.3.3: Example to read register 2090. Physical unit is set to mV (0x200000), PMC1 is

175.9922 mV, Status is 0x00, Min allowed value is -414.0028 mV, Max allowed value is
414.0028 mV.

For the definition of the measurement status see chapter 2.5.4.

Attention:

You cannot read selectively the registers 3 and 4 for the measurement value only. You have to read the
entire length of the command (10 registers) and extract the desired information.