pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 58 / 67

2.8.2 Operating Hours and Counters

In register 4676 are given:

 total operating hours

 operating hours above max measurement temperature (see chapter 2.8.1)

 the operating hours above max operating temperature (see chapter 2.8.1)

In register 4682 are stored:

 number of power ups

 number of watchdog resets

 number of writing cycles to the sensor’

s flash memory

In register 4688 are given:

 number of sterilizations in place (SIP) (see chapter 2.8.5)

 number of cleanings in place (CIP) (see chapter 2.8.5)

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

4676

6

Operating
hours
[h]

Operating
hours above
max
measurement
temperature
[h]

Operating
hours above
max
operating
temperature
[h]

3, 4

U/A/S

none

4682

6

Number of
Power ups

Number of
Watchdog
resets

Number of
Writing
cycles to
flash memory

3, 4

U/A/S

none

4688

4

Number of
SIP cycles

Number of
CIP cycles

-

3, 4

U/A/S

none

Figure 2.8.2.1: Definition of register 4676, 4682 and 4688.

Command: Operating hours

Modbus address:

4676

Length:

6

Type:

3

Read

Parameter:

Operating hours [h]

Operating hours
above max
measurement
temperature
[h]

Operating hours
above max
operating
temperature
[h]

Format:

Float

Float

Float

Value:

168.3667

0

0

Figure 2.8.2.2: Example to read the total operating hours, the operating hours above the max

measurement temperature and the operating hours above the max operating temperature.

Command: Power & watchdog

Modbus address:

4682

Length:

6

Type:

3

Read

Parameter:

Number of Power
ups

Number of
Watchdog resets

Number of Writing
cycles to flash
memory

Format:

Decimal

Decimal

Decimal

Value:

34

1

16

Figure 2.8.2.3: Example to read the number of power ups, the number of watchdog resets and the

number of writing cycles to flash memory.

Command: SIP & CIP

Modbus address:

4688

Length:

4

Type:

3

Read

Parameter:

SIP cycles

CIP cycles

Format:

Decimal

Decimal

Value:

0

0

Figure 2.8.2.4: Example to read the number of SIP cycles and the number of CIP cycles. For the

definition of SIP and CIP cycles see chapter 2.8.5.