pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 7 / 67

2.3

Configuration of the serial RS485 Interface

Factory settings of the RS485:
Parity is none, 1 start bit, 8 data bits, 2 stop bits (in total: 11 bits).

2.3.1 Device Address

2.3.1.1

Reading and Writing the Device Address

The sensor specific device address can be read and written in register 4096.

Start

register

Number of
registers

Reg1 / Reg2

Modbus
function code

Read
access

Write
access

4096

2

device address

3, 4, 16

U/A/S

S

Figure 2.3.1.1.1: Definition of register 4096.

Command: Com address

Modbus address:

4096

Length:

2

Type:

3

Read

Parameter:

Modbus address

Format:

Decimal

Value:

1

Figure 2.3.1.1.2: Example to read the device address.

The device address can be set by S (Specialist), default value is 1.

Command: Com address

Modbus address:

4096

Length:

2

Type:

16

Write

Parameter:

Modbus address

Format:

Decimal

Value:

3

Figure 2.3.1.1.3: Example to set the device address to 3.

2.3.1.2

Reading the Device Address Limits

The device address limits can be read in register 4098.

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Modbus
function code

Read
access

Write
access

4098

4

Min. device
address

Max. device
address

3, 4

U/A/S

none

Figure 2.3.1.2.1: Definition of register 4098.

Command: Com address limits

Modbus address:

4098

Length:

4

Type:

3

Read

Parameter:

Min value

Max value

Format:

Decimal

Decimal

Value:

1

32

Figure 2.3.1.2.2: Example to read the device address limits: Min = 1, Max = 32.