6 configuration of the measurement – Hamilton Electronics EPHUM011 User Manual

Page 26

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 26 / 67

2.6

Configuration of the Measurement

This chapter describes the configuration of PMC1 and PMC6 by means of measurement parameters
(PA).

2.6.1 Available Parameters

In register 3072, all available parameters (PA) are given.

Start
register

Number of
registers

Reg1 / Reg2
(bitwise defined)

Modbus
function code

Read
access

Write
access

3072

2

Available parameters (see figure 2.6.1.2)

3, 4

U/A/S

none

Figure 2.6.1.1: Definition of register 3072.

Bit #

Hex value

Description

Definition in pH ARC Sensors

0 (LSB)

0x0001

PA1

not available
not available

7

0x0080

PA8

not available

8

0x0100

PA9

Moving average

9

0x0200

PA10

not available

10

0x0400

PA11

not available

11

0x0800

PA12

Moving average R
not available

15 (MSB)

0x8000

PA16

not available

Figure 2.6.1.2: Bitwise definition of parameters PA1 to PA16, valid for pH ARC Sensors

Command: Available parameters

Modbus address:

3072

Length:

2

Type:

3

Read

Parameter:

Measurement
parameters

Format:

Hex

Value:

0x0900

Figure 2.6.1.3: Example to read the available parameters. The value 0x0900 corresponds to 0x0100

(PA9) + 0x0800 (PA12). Parameter 9 and 12 are available.

Genereal note:

 PA1 to PA8 use FLOAT as data format for its values

 PA9 to PA16 use UNSIGNED INT as data format for its values.