Hamilton Electronics EPHUM011 User Manual

Page 33

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 33 / 67

2.7.2 Definitions of Calibration Points

2.7.2.1

Calibration Points 1 and 2 (Standard Calibration)

The limits for the calibration point 1 are defined in register 5152, for calibration point 2 in register 5184.

Start
register

Number of
registers

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus
function code

Read
access

Write
access

5152

6

Physical unit
currently
active for
CP1

Min value for
CP1 (in the
physical unit
as defined in
Reg1 and 2)

Max value for
CP1 (in the
physical unit
as defined in
Reg1 and 2)

3, 4

U/A/S

none

5184

6

Physical unit
currently
active for
CP2

Min value for
CP2 (in the
physical unit
as defined in
Reg1 and 2)

Max value for
CP2 (in the
physical unit
as defined in
Reg1 and 2)

3, 4

U/A/S

none

Figure 2.7.2.1.1: Definition of register 5152 for CP1 and 5184 for CP2.

Attention:

The only physical unit available for calibration is pH ! The physical unit defined in 5152, 5184 and 5312
for CP1, CP2 and CP6 is NOT linked to the physical unit defined for PMC1 in register 2090.
When performing a calibration while having the physical unit set to

mV

, the calibration status will

report

CP1: incorrect measurement unit

”or “

CP2: incorrect measurement unit

”(see figure 2.7.4.1.1).

Command: Calibration limits CP1

Modbus address:

5152

Length:

6

Type:

3

Read

Parameter:

Unit

Min value

Max value

Format:

Hex

Float

Float

Value:

0x01000

0

0

Figure 2.7.2.1.2: Example to read the limits of CP1. Currently active physical unit is pH (0x01000). The

min and max values are both 0, indicating, that calibration at CP1 can be performed only using
defined calibration standards having discrete pH values.

Command: Calibration limits CP2

Modbus address:

5184

Length:

6

Type:

3

Read

Parameter:

Unit

Min value

Max value

Format:

Hex

Float

Float

Value:

0x01000

0

0

Figure 2.7.2.1.3: Example to read the limits of CP2. The active physical unit is is pH (0x01000). The

min and max values are both 0, indicating, that calibration at CP2 can be performed only using
defined calibration standards having discrete pH values.

When initiating the calibration at CP1 and CP2, the measured pH and temperature have to be stable for
at least 3 minutes. The stability criteria are defined in register 5128:

Start
register

Number of
registers

Reg1 / Reg2
(Float)

Reg3 / Reg4
(Float)

Modbus
function code

Read
access

Write
access

5128

4

Max. Drift PMC1
pH
[pH/min]

Max. Drift PMC6
Temperature
[K/min]

3, 4, 16

U/A/S

S

Figure 2.7.2.1.4: Definition of register 5128.