2.8.4 Errors

An

Error

”message indicates a serious problem of the sensor which does not allow further proper

functioning of the sensor. This problem must be solved.

2.8.4.1

Currently Active Errors

The currently active errors are stored in register 4800.

Start
register

Number
of
registers

Reg1 /
Reg2

Reg3 /
Reg4

Reg5 /
Reg6

Reg7 /
Reg8

Modbus
function
code

Read
access

Write
access

4800

8

Active error
measure-
ment
(bitwise
defined)

Active error
calibration
(bitwise
defined)

Active error
interface
(bitwise
defined)

Active error
hardware
(bitwise
defined)

3, 4

U/A/S

none

Figure 2.8.4.1.1: Definition of register 4800 (see chapter 2.8.4.3)

Command: Active errors

Modbus address:

4800

Length:

8

Type:

3

Read

Parameter:

E Measurement

E Calibration

E Interface

E Hardware

Format:

Hex

Hex

Hex

Hex

Value:

0x00

0x00

0x00

0x00

Figure 2.8.4.1.2: Example to read the currently active errors.

2.8.4.2

History of Errors

The history of errors is not implemented in pH ARC Sensors.