Hamilton Electronics EPHUM011 User Manual

Page 61

Advertising
background image

pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 61 / 67

2.8.4.3

Definition of Errors

Bit #

Hex

Description
not available

5

0x0000020

Glass resistance too high

6

0x0000040

Glass resistance too low

7

0x0000080

Reference electrode resistance too high

8

0x0000100

Reference electrode resistance too low
not available

15

0x0008000

Auxiliary electrode electrical potential too high

16

0x0010000

Auxiliary electrode electrical potential too low

17

0x0020000

Auxiliary electrode resistance too high

18

0x0040000

Auxiliary electrode resistance too low
not available

25

0x2000000

Temperature sensor defective

Figure 2.8.4.3.1: Definition of errors

measurement

.

Bit #

Hex

Description

1

0x0000002

Sensor failure (Quality value < 15%)

Figure 2.8.4.3.2: Definition of errors

calibration

.

Bit #

Hex

Description
not available

Figure 2.8.4.3.3: Definition of errors

interface

. None is defined.

Bit #

Hex

Description
not available

24

0x1000000

Internal communication error (between front-end and user-end)

Figure 2.8.4.3.4: Definition of errors

hardware

.