pH ARC Sensors Modbus RTU Programmer

s Manual (EPHUM011)

624300/01

page 38 / 67

Examples: (Definitions of register 5158 and 5190 used in these examples are given in chapter 2.7.4.1,
those for register 4872 in chapter 2.8.6)

Example to calibrate at CP1 with automatic recognition:

Command: Make calibration CP1

Modbus address:

5162

Length:

2

Type:

16

Write

Parameter:

pH value

Format:

Float

Value:

0

Figure 2.7.3.1.4: Example to start the calibration at CP1, setting a value of 0 for automatic

recognition of the calibration standard.

Example to calibrate at CP2 with automatic recognition:

Command: Make calibration CP2

Modbus address:

5194

Length:

2

Type:

16

Write

Parameter:

pH value

Format:

Float

Value:

0

Figure 2.7.3.1.5: Example to start the calibration at CP2, setting a value of 0 for automatic

recognition of the calibration standard.

Example to read the calibration status of CP1:

Command: Calibration status CP1

Modbus address:

5158

Length:

6

Type:

3

Read

Parameter:

Status

Unit

Value

Format:

Hex

Hex

Float

Value:

0x00000000

0x00001000

4.01

Figure 2.7.3.1.6: Example to read the calibration status of CP1 after calibration CP1 at 0 = auto.

All bits of CP1 are zero (0x000000

00), indicating that the calibration was successful. The

physical unit of the last calibration is pH (0x00001000) and the assigned pH value is 4.01
at 25°C.

Example to read the calibration status of CP2:

Command: Calibration status CP2

Modbus address:

5190

Length:

6

Type:

3

Read

Parameter:

Status

Unit

Value

Format:

Hex

Hex

Float

Value:

0x00000000

0x00001000

7

Figure 2.7.3.1.7: Example to read the calibration status of CP2 after calibrating CP2 at 0 = auto.

All bits of CP2 are zero (0x0000

0000), indicating that the calibration was successful. The

physical unit of the last calibration is pH (0x00001000) and the pH value is 7.

Example to calibrate at CP1 with manual selection of the calibration standard:

Command: Make calibration CP1

Modbus address:

5162

Length:

2

Type:

16

Write

Parameter:

pH value

Format:

Float

Value:

4.00

Figure 2.7.3.1.8: Example to start the calibration at CP1, by means of manually selecting the

calibration standard HAMILTON DURACAL 4.01 (nominal value 4.01). The operator knows
from the certificate of the specific production lot that the actual pH value is 4.00.