3 distributed mounting of miq modules, Distributed mounting of miq modules -21, Section 3.5.3 – YSI IQ SensorNet 2020 XT Terminal User Manual

Page 51

Advertising
background image

System 2020 XT USB

Installation

3 - 21

ba76019e01

12/2012

3.5.3

Distributed mounting of MIQ modules

General information

The following IQ S

ENSOR

N

ET

cables can be used for distributed

mounting:

 SNCIQ cable

 SNCIQ/UG earth cable - suitable for underground laying according

to VDE 01816, Part 2 and DIN/VDE 0891, Part 6.

The cables are delivered as piece goods (please specify length when
ordering!).

Each MIQ module has a row of terminal strips in the housing. The two
3-fold terminal strips on the right are always reserved for the
connection of further IQ S

ENSOR

N

ET

components (designation

"SENSORNET 1" and "SENSORNET 2").

Some MIQ modules have additional "SENSORNET" connections
("SENSORNET 3" and "SENSORNET 4") on the left of the terminal
strip.

Caution

The IQ S

ENSOR

N

ET

cable may only be connected to the

SENSORNET connections. No wire of the cable may be connected
with an external electrical potential. Otherwise, malfunctions
could occur.

Note
IQ S

ENSOR

N

ET

cables must always be installed separately at a

minimum distance of 20 cm from other lines that carry a voltage greater
than 60 V.

Materials required

 1 x SNCIQ or SNCIQ/UG connection cable (see chapter 10

A

CCESSORIES

AND

OPTIONS

)

 Wire end sleeves for 0.75 mm

2

wire cross-section with matching

crimping tool

 1 x cable gland with seal (scope of delivery of MIQ module).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screw driver

 Small screw driver.

9

Close the back MIQ module and fix it with the two countersunk
screws (pos. 5 in Fig. 3-13).