112

Figure 67 - Language Selection Screen

Self Calibration

Self Calibration procedure can optimize the signal path for optimal

measurement results. You can run this procedure at any time. If the

operating temperature changes by more than 5° C or the instrument runs

more than thirty minutes, running self calibration is recommended.

When you run the self Calibration, you must disconnect all the probes

and leads, and make sure the scope is not in Single-Shot mode. Then

press the “Utility” button to choose the “Do self cal” to show the self

Calibration Menu, and execute self calibration program according to

the prompt on the screen.